Conductivity measures how well water conducts electricity because of dissolved ions. Resistivity is the inverse. pH shows the level of acidity or alkalinity. You must read all three at the same reference temperature. A single number cannot prove water safety or identify a specific contaminant.
PT Watermart Perkasa supplies water-treatment instruments and components in Indonesia. In RO, demineralization, process water and drinking water operations, conductivity, resistivity and pH data are most useful as trends against an accepted baseline. They are not a standalone quality label.
| Parameter | What is actually measured | Common units | Operational decisions supported |
|---|---|---|---|
| Conductivity | The solution’s ability to conduct current due to dissolved ions | µS/cm or mS/cm | Detecting changes in raw water, RO salt passage, ion leakage, or process concentration |
| Resistivity | The solution’s electrical resistance; the inverse of conductivity | MΩ·cm | Monitoring deionized water or mixed beds in the very low range |
| pH | Hydrogen ion activity on a logarithmic scale | pH units | Controlling corrosion, coagulation, disinfection, and process suitability |
At the same temperature and in the same units, the practical relationship is resistivity (MΩ·cm) = 1 / conductivity (µS/cm). For example, 0.10 µS/cm is equivalent to 10 MΩ·cm. Do not convert readings recorded at different temperatures without compensation, because conductivity changes with temperature.
How to Read Conductivity, Resistivity, and pH Together
Start from the accepted raw water and product water baseline. Then compare changes in the three parameters at the same sampling point and reference temperature. The following matrix helps determine the next checks. The results are not a laboratory diagnosis.
| Data pattern | Possible causes to check | Verification action |
|---|---|---|
| Conductivity rises, pH relatively stable | Increased ion load in raw water, RO salt passage, or mixing leakage | Repeat the sample, check temperature, compare feed-permeate, then check RO pressure and flow |
| Resistivity falls slowly | Resin approaching exhaustion, loop contamination, or declining upstream RO quality | Check bed volume trend, rinse, and samples before and after the mixed bed |
| pH changes, conductivity almost unchanged | Acid/base dosing, CO₂ absorption, or electrode problem | Recalibrate the pH meter and take a closed sample if CO₂ is relevant |
| All values change suddenly | Wrong sampling point, cross-contamination, temperature change, or process upset | Hold the decision, verify the instrument with a standard, then resample |
ISO 7888:1985, confirmed by ISO in 2023 as still valid, specifies the method for measuring conductivity in surface water, process water, treatment plants and wastewater. This standard states that conductivity can be used for monitoring. However, its interpretation must take measurement interference and the purpose of the analysis into account.
What Is Conductivity in Water Treatment?
Conductivity is an indicator of the relative amount of ionic species moving in water. A low value means low ionic conduction. However, it does not prove the water is free of microorganisms, neutral organic compounds or particles. The final product specification must still be determined by the applicable process and regulations.
In reverse osmosis (RO), compare feed and permeate conductivity under stable operating conditions to calculate salt rejection: rejection (%) = (1 − permeate conductivity / feed conductivity) × 100. Conductivity is used as a TDS approximation, so this figure is an estimate of salt rejection. It is the complement of salt passage, that is, the percentage of salt that passes into the permeate. It is not a direct TDS measurement. This trend can indicate performance changes. However, it must be read together with pressure, temperature, recovery and flow. Low feed conductivity alone does not prove the membrane is working more effectively.

Why Resistivity Matters?
Resistivity matters in demineralized and deionized water. In this range, small changes in ion content are easier to see. A high value means high electrical resistance. However, you must not assign quality labels such as pharmaceutical or semiconductor from resistivity alone. Each user needs to apply their own specifications and verification methods.
Periodic resistivity measurement also helps to monitor the performance of water treatment systems, such as ion exchange resin or filtration membranes. A sudden increase in mineral contamination shows as a drop in resistivity. This can be an early sign that system components need maintenance or replacement. Early detection matters for water quality and for keeping operating costs down. It also keeps production downtime low.

pH and Its Effect on Water Quality
pH adjustment affects corrosion, coagulation, carbonate equilibrium and the chemical form of disinfectants. pH is on a logarithmic scale. The electrode is sensitive to temperature, junction condition and the storage liquid. Therefore operators need to record the calibration results together with the process values, and not just the number on the screen.
In the food and beverage industry, pH can affect taste, formulation stability, corrosion and how well certain process steps work. However, a suitable pH does not prove microbiological safety. That parameter still needs its own test methods and acceptance limits.
Temperature compensation, automatic calibration and digital output can reduce handling errors. However, they do not replace verification with standards, electrode maintenance and recording of calibration results.
The pH target must come from process requirements or product water specifications. Alarms should have warning limits, action limits and verification procedures. Then a single deviating reading does not immediately trigger chemical overdosing.

Conductivity, resistivity and pH give fast process signals. However, the three are not a substitute for contaminant or microbiological analysis. They work best when instruments are calibrated, sample points are consistent, temperature is recorded and trends are tied to clear operator action.
Measurement and Instrument Handover Checklist
- Set the sample point, working range, reference temperature and alarm limits from the design data or the accepted baseline.
- Choose a suitable cell constant and instrument range. Do not use a high-range sensor for ultrapure water without evidence of adequate resolution.
- Check that the calibration standards are still valid and cover the measuring range. Record the lot number, the temperature and the results before and after calibration.
- Rinse the probe and container with the sample. Avoid bubbles, wait for a stable reading, then record the temperature together with the result.
- For pH, calibrate at two or three points according to the process range. Check the slope and offset as described in the instrument manual.
- During commissioning, compare the online instrument with a portable reference meter on the same sample. Test the alarm or 4–20 mA output.
- Keep trends for feed, permeate and product. Look into process changes before you change a setpoint or replace media.
For online monitoring, see industrial conductivity monitors and flowmeters. If trends show salt passage or a drop in purity, evaluate the reverse osmosis application and mixed bed polishing along with the condition of the resin, membranes, valves and sample points. Drinking water compliance must still be proven against the applicable health parameters. Permenkes No. 2 Tahun 2023 is the Indonesian environmental health regulation reference in force for this update.
Data for Selecting a Conductivity Analyzer
When you request a quotation for an analyzer, do not send only a single target µS/cm value. Sensor range, materials, temperature compensation, installation and control output must match the process conditions and the existing panel.
| Data submitted | Example of the information needed | Effect on selection |
|---|---|---|
| Measuring point and function | RO feed, permeate, mixed bed outlet, or product alarm | Determines the range, cell constant, and detection limit |
| Minimum, normal, and maximum values | Historical data together with sample temperature | Prevents a sensor that is too coarse or falls out of range quickly |
| Process conditions | Temperature, pressure, flow, pipe material, and potential chemicals | Determines wetted parts material and installation method |
| Panel requirements | 4–20 mA, relay, communication, power supply, and enclosure | Determines the transmitter and control integration |
| Verification | Calibration standards, comparison instrument, and inspection interval | Determines sensor access and the commissioning plan |
Attach photos of the installation point, connection sizes, distance to the panel and examples of actual trends. Complete data lets PT Watermart Perkasa tell the sensor requirements of an ordinary process apart from the measurement of very low-conductivity water.
Buyer Questions about Water Conductivity
Is conductivity the same as TDS?
No. Conductivity is a direct electrical reading. Many instruments display TDS as an estimate from conductivity using a conversion factor. That factor depends on ion composition. Therefore TDS obtained by conversion needs to be distinguished from TDS determined by a laboratory method.
What is a good conductivity for water?
No single number suits every use. Set the limit from the product water specification, the accepted baseline and the next process. Drinking water, RO feed, RO permeate, boiler water and deionized water need different ranges and verification parameters.
When does a conductivity sensor need to be calibrated?
Follow the manufacturer’s interval and the quality system. Then check again after cleaning, after you replace the sensor, after improper storage, or when the reading differs from the reference instrument. Use standards that are still valid and suitable for the working range.